CalibKeepr ยท Metrology Specifications
What the Calibrations Need to Be (SOPs & Tolerances)
Define the exact step-by-step checklist, target nominal values, +/- tolerances, ambient temperature/humidity limits, and master reference standards required for every tool family.
Digital & Dial Caliper 4-Point Dimensional Calibration (0–150 mm)
Verifies zero repeatability, jaw parallelism, outside diameter accuracy, and inside jaw offset against Grade 0 Gage Blocks.
Step-by-Step Procedure Checklist (5 steps)
Clean external/internal jaws and beam slide with IPA. Place caliper and Grade 0 gage blocks on granite plate for 60 minutes.
Close outside jaws gently and hold against a uniform light source to inspect for light leakage along the contact length.
Zero the digital display with jaws closed, open to 50 mm and close 5 times consecutively.
Measure the 25.00 mm, 75.00 mm, and 150.00 mm Grade 0 gage blocks at the center of the outside jaws.
Affix tamper-evident calibration sticker with Certificate ID, Technician Initials, and Next Due Date.
Target Measurement Checkpoints & Tolerances (4 points)
| # | Checkpoint Parameter | Nominal Target | Tolerance | Acceptable Range | |
|---|---|---|---|---|---|
| 1 | Closed Outside Jaws (Zero) | 0 mm | -0.01 / +0.01 mm | -0.01 to 0.01 mm | |
| 2 | Outside Jaw — 25 mm Gage Block | 25 mm | -0.02 / +0.02 mm | 24.98 to 25.02 mm | |
| 3 | Outside Jaw — 75 mm Gage Block | 75 mm | -0.02 / +0.02 mm | 74.98 to 75.02 mm | |
| 4 | Outside Jaw — 150 mm Gage Block | 150 mm | -0.02 / +0.02 mm | 149.98 to 150.02 mm |
Electronic & Click Torque Wrench Verification (20%, 60%, 100% Scale)
Three-point clockwise torque verification at 20%, 60%, and 100% of maximum nominal torque capacity on the master torque transducer bench.
Step-by-Step Procedure Checklist (5 steps)
Inspect square drive, ratchet pawl, and battery/display. Cycle wrench 5 times at 80 Nm on transducer bench.
Verify ambient temperature (18–24°C) and zero the CDI 5000-ST transducer in Track/Peak mode with no load applied.
Set wrench to 20.00 Nm. Apply slow, steady clockwise load (1–2 sec to peak) and record peak transducer reading.
Repeat steady clockwise verification at 60.00 Nm and 100.00 Nm settings.
Return mechanical scale to minimum setting (if click-type), reset station cycle counter, and attach calibration tag.
Target Measurement Checkpoints & Tolerances (3 points)
| # | Checkpoint Parameter | Nominal Target | Tolerance | Acceptable Range | |
|---|---|---|---|---|---|
| 1 | Low Range (20% Scale) | 20 Nm | -0.8 / +0.8 Nm | 19.2 to 20.8 Nm | |
| 2 | Mid Range (60% Scale) | 60 Nm | -2.4 / +2.4 Nm | 57.6 to 62.4 Nm | |
| 3 | Full Scale (100% Scale) | 100 Nm | -4 / +4 Nm | 96 to 104 Nm |
Outside Micrometer Optical Flat & Gage Block Calibration (0–25 mm)
High-precision verification of anvil/spindle flatness using monochromatic optical flats and lead-screw linearity using staggered gage block sizes.
Step-by-Step Procedure Checklist (3 steps)
Draw clean bond paper between lightly closed anvil and spindle faces. Stabilize in Metrology Lab for 120 minutes.
Place optical flat on anvil and spindle under monochromatic helium light and count interference fringes.
Measure Grade 0 blocks at 6.500 mm, 12.500 mm, and 25.000 mm to test spindle pitch at fractional rotations.
Target Measurement Checkpoints & Tolerances (4 points)
| # | Checkpoint Parameter | Nominal Target | Tolerance | Acceptable Range | |
|---|---|---|---|---|---|
| 1 | Anvil Contact Zero | 0 mm | -0.001 / +0.001 mm | -0.001 to 0.001 mm | |
| 2 | 6.500 mm Fractional Rotation | 6.5 mm | -0.002 / +0.002 mm | 6.498 to 6.502 mm | |
| 3 | 12.500 mm Half-Span Block | 12.5 mm | -0.002 / +0.002 mm | 12.498 to 12.502 mm | |
| 4 | 25.000 mm Full-Span Block | 25 mm | -0.002 / +0.002 mm | 24.998 to 25.002 mm |
Hydraulic Digital Pressure Gauge 4-Point Rising/Falling Calibration (0–500 PSI)
Compares manifold gauge readings against Fluke 754 / 750P06 Reference Module across 0%, 25%, 50%, and 100% span.
Step-by-Step Procedure Checklist (2 steps)
Follow LOTO to isolate gauge, bleed residual hydraulic pressure, and mount on comparison test pump.
Apply 0.0, 125.0, 250.0, and 500.0 PSI on rising and falling cycles to verify hysteresis and linearity.
Target Measurement Checkpoints & Tolerances (4 points)
| # | Checkpoint Parameter | Nominal Target | Tolerance | Acceptable Range | |
|---|---|---|---|---|---|
| 1 | Ambient Vented Zero (0% FS) | 0 PSI | -0.5 / +0.5 PSI | -0.5 to 0.5 PSI | |
| 2 | Quarter Span (25% FS) | 125 PSI | -1.5 / +1.5 PSI | 123.5 to 126.5 PSI | |
| 3 | Mid Span (50% FS) | 250 PSI | -1.5 / +1.5 PSI | 248.5 to 251.5 PSI | |
| 4 | Full Scale Proof (100% FS) | 500 PSI | -1.5 / +1.5 PSI | 498.5 to 501.5 PSI |
Curing Oven RTD & Infrared Pyrometer Dry-Block Calibration (50–250 °C)
Three-point thermal comparison in dry-well bath against NIST-traceable secondary PRT probe.
Step-by-Step Procedure Checklist (2 steps)
Place Fluke 5615 PRT and sensor under test into matching sleeves at full immersion depth.
Stabilize 10 minutes per setpoint and record Reference PRT vs. Tool reading.
Target Measurement Checkpoints & Tolerances (3 points)
| # | Checkpoint Parameter | Nominal Target | Tolerance | Acceptable Range | |
|---|---|---|---|---|---|
| 1 | Low Setpoint (50 °C) | 50 °C | -0.5 / +0.5 °C | 49.5 to 50.5 °C | |
| 2 | Cure Soak Setpoint (150 °C) | 150 °C | -0.8 / +0.8 °C | 149.2 to 150.8 °C | |
| 3 | High Limit Setpoint (250 °C) | 250 °C | -1 / +1 °C | 249 to 251 °C |