CalibKeepr Β· Automated Alerts & Calendar Feed
Calibration Reminders, Digest & Calendar Sync
Configure how far in advance the factory is reminded of upcoming calibrations, set usage cycle warning thresholds, preview or dispatch HTML email digests, and sync with Outlook / Google Calendar.
REMINDER RULES & EMAIL DIGEST
Plant Notification Preferences
Applies to the top notification bell, dashboard urgency banners, and the background email digest worker.
Tools due within this window appear as "Due Soon".
Triggers early usage warning before reaching 100% cycle limit.
Last digest sent: Oct 9, 2026 4:32 PM
EXTERNAL CALENDAR SYNC (.ICS)
Subscribe in Outlook, Google, or Apple Calendar
Every active tool calibration due date and master reference standard expiration date is published to this live iCalendar feedβcomplete with required SOP codes and ambient specs in the event description.
Active Reminder Triggers (7 items)
Overdue Calibration · Final Assembly Line 2
TRQ-204: Snap-on TechAngle Electronic Torque Wrench (10–100 Nm)
Overdue by 5 day(s) (Due Oct 4, 2026).
Overdue Calibration · Heat Treat & Hydraulic Test Bay
PRG-302: Ashcroft Digital Test Pressure Gauge (0–500 PSI)
Overdue by 3 day(s) (Due Oct 6, 2026).
Out-of-Tolerance Quarantine · Heat Treat & Hydraulic Test Bay
TMP-505: Omega Platinum RTD Curing Oven Probe #1 (50–250 °C) is Quarantined
Failed 250.0 °C checkpoint (+1.65 °C drift vs ±1.00 °C limit). Tagged RED Out-of-Tolerance pending replacement.
Cycle-Count Trigger · Final Assembly Line 2
TRQ-209: Desoutter DC Electric Nutrunner Spindle #2 (10–100 Nm)
2,640 / 2,500 cycles used (106%).
Master Reference Standard
STD-TRQ-02: CDI Suretest 5000-ST Master Torque Transducer Bench expires in 18 days